Hach DR900 Multiparameter Colorimeter: Complete Review & Reference (2026)

Source: Hach DR 900 User Manual DOC022.97.80344, Edition 4, April 2022 • DR900 Methods Reference DR900_CMS_block.pdf

Product No. 9385100 • 90+ Methods • 48+ Parameters • 420 / 520 / 560 / 610 nm • IP67 • 500 Results Storage

Hach DR900 Multiparameter Portable Colorimeter showing settings menu with Favorites, All Programs, Data Log, and Setup options Water treatment operator using Hach DR900 Multiparameter Colorimeter at a wastewater treatment facility
Hach DR900 Multiparameter Portable Colorimeter — Product No. 9385100
90+ methods • 48+ parameters • 4 wavelengths • IP67 waterproof • 500-result data log • USB data export. Reagents sold separately.
Buy on Amazon →
Product No.
9385100
Methods
90+ pre-programmed
Parameters
48+
Wavelengths
420 / 520 / 560 / 610 nm
Light Source
LED (4 wavelengths)
Data Storage
500 results
IP Rating
IP67 Waterproof
Power
4× AA alkaline batteries
Connectivity
USB mini (data export)
User Programs
Up to 10 (favorites + custom)
Languages
9 (incl. EN, ES, FR, ZH, AR)
Manual
DOC022.97.80344 Ed.4 (2022)

DR900 vs DR300 — Which Instrument Do You Need?

The single most important thing to understand about the DR900: it is a fundamentally different class of instrument from the single-parameter DR300 Pocket Colorimeters. The DR300 family (chlorine, iron, phosphate, bromine) is built for one parameter per instrument at one fixed wavelength. The DR900 measures 48+ parameters across four wavelengths from a single device. That difference drives every purchasing decision.

DR300 vs DR900 — DECISION GUIDE
DR300 Pocket Colorimeter
✓ Single parameter per unit
✓ Lower instrument cost
✓ Simpler operation — 2 buttons
✓ Compact — fits in a shirt pocket
✓ IP67 waterproof
✗ No data logging or USB export
✗ One instrument per parameter
✗ No Standard Adjust or user programs

Best for: routine monitoring of a single parameter at one site; low budget; minimal data management needs.
DR900 Multiparameter
✓ 48+ parameters, one instrument
✓ 500-result data log + USB export
✓ Standard Adjust calibration
✓ Operator & Sample ID tagging
✓ User-programmable methods
✓ Favorites in <4 key presses
✓ IP67 + drop-tested
✗ Higher instrument cost

Best for: multi-parameter monitoring; compliance reporting with data records; field operators covering multiple sites or test types.
FeatureDR300 (single-param)DR900 (multiparameter)
Parameters per instrument148+
Pre-programmed methods190+
Wavelengths1 (fixed)4 (420, 520, 560, 610 nm)
Data loggingNone500 results with timestamp
USB connectivityNoneUSB mini — CSV export
Operator/Sample IDNoneYes (10 IDs each)
Standard AdjustSCA onlyFull Standard Adjust
User programsNoneUp to 10 (custom + favorites)
Sample cellsModel-specific16 mm vial, 25 mm round, 1 cm plastic
IP ratingIP67IP67 + drop-tested
Batteries4× AAA4× AA

Complete Technical Specifications

SpecificationValue
Product number9385100
Light sourceLED at 420, 520, 560, and 610 nm
Wavelength selectionAutomatic based on selected program
Pre-programmed methods90+ methods
Parameters covered48+
User programs / favoritesUp to 10 combined
Data storage500 results (data log)
Data exportUSB mini — DR900_DataLog.csv and DR900_EventLog.csv
Operator IDsUp to 10
Sample IDsUp to 10
DisplayBacklit LCD
Keypad6 keys: POWER, HOME, SETTINGS, BACKLIGHT, navigation d-pad, 2 soft keys
Sample cells accepted16-mm glass test vials (with adapter); 25 mm (1-inch) round glass cells; 1 cm / 10 mL plastic cells
IP ratingIP67 waterproof (submersible to 1 m for 30 min)
Drop resistanceDrop-tested per IEC 68-2-32
Power source4× AA alkaline batteries
LanguagesEnglish, French, Spanish, Portuguese, Chinese (Simplified), Japanese, Korean, Thai, Arabic
ManualDOC022.97.80344, Edition 4 (April 2022)

Sample Cells — Three Types, One Compartment

The DR900 accommodates all three Hach sample cell formats through a single cell compartment using a removable cell adapter. Using the wrong cell type or omitting/including the adapter incorrectly will produce inaccurate results — the optical path length changes with cell diameter.

Cell TypeSizeAdapter RequiredCommon Use
16-mm glass test vial16 mm diameterYes — adapter must be installedMost Hach reagent kit methods; the most common cell type across the 90+ methods
25 mm round glass cell1-inch (25 mm) diameterNo — insert directlyDR300-compatible methods (chlorine DPD, iron FerroVer, phosphate); large sample volume methods
1 cm / 10 mL plastic cell1 cm squareNo — insert directlySelect methods; avoid scratched or cloudy cells
Cell adapter orientation matters. When using 16-mm vials, install the cell adapter with the correct orientation for the vial to seat in the optical path. Refer to Figure 3 in the user manual for the correct adapter position. Wipe vial exteriors with a lint-free cloth before inserting — fingerprints on the optical surface affect absorbance readings.

Keypad — 6-Key Interface Reference

Key / ControlFunction
POWERTurn instrument on or off. Hold to power off from any screen.
HOMEReturn to the main reading screen from any menu.
SETTINGSAccess the Settings menu: Setup (language, date/time), Operator ID, Sample ID, and instrument configuration.
BACKLIGHTToggle the display backlight on and off for use in bright outdoor conditions or to extend battery life.
Navigation d-pad (4-way)UP/DOWN to scroll through program lists, menus, and numeric values. LEFT/RIGHT to navigate between fields. Confirms selections when used as directed by screen prompts.
Soft keys (2×)Context-sensitive. Label changes based on the active screen. Most common labels: Options, Select, Start, OK, Done, Back.

Selecting and Running a Program — Standard Workflow

1

Power on the instrument

Press POWER. The main reading screen displays the last used program. Battery level appears in the upper-right corner.

2

Select a program

Press HOME → Options → All Programs. Scroll alphabetically or numerically to find the method. Highlight and press Start (right soft key). For frequently used tests, add to Favorites first (see below) for access in fewer than 4 key presses.

3

Assign Operator ID and Sample ID (if required)

Press SETTINGS → Operator ID to select a pre-entered ID. Press SETTINGS → Sample ID to assign a sample location. Both tags appear in the data log entry for this reading.

4

Prepare the blank cell

Fill the appropriate sample cell with the sample water (no reagent). Wipe the exterior with a lint-free cloth. Install the cell in the instrument (with or without adapter depending on cell type). Press the Zero soft key. The instrument subtracts the sample’s natural absorbance from the subsequent reading.

5

Prepare the sample cell with reagent

Follow the specific method procedure document for the selected parameter — add the correct reagent type and quantity, observe any mandatory reaction time, and swirl or mix as directed. Wipe exterior clean before inserting.

6

Insert sample cell — press Read

Insert the reagent-containing sample cell into the compartment. Confirm correct adapter status for the cell type. Close the compartment lid if applicable. Press Read (right soft key). Result displays in the configured unit.

7

Reading is automatically stored

Every reading is automatically stored to the data log (500-result capacity, FIFO). The entry includes parameter, value, unit, program number, Operator ID, Sample ID, and date/time stamp.

8

Interpret the “+++” and “---” over-range indicators

“+++” means the measured value exceeds the upper limit of the method range — dilute the sample with reagent-demand-free water and multiply the result by the dilution factor. “---” means the value is below the lower detection limit of the method — verify the test procedure and reagent quality.

Favorites — Access Any Method in Under 4 Key Presses

Add to Favorites immediately after finding a method. The Favorites list is the operational efficiency feature most operators underuse. Once a method is in Favorites, it appears at the top of the program selection screen — no scrolling through 90+ programs. Maximum 10 programs across Favorites and User Programs combined.
1

Select any program from All Programs

Navigate to HOME → Options → All Programs. Highlight the target method.

2

Press Options → Add to Favorites

The method is saved to the Favorites list and will appear at the top of the HOME → Options screen on future power-ons.

Reagent Blank Correction — Step by Step

The reagent blank correction removes background absorbance contributed by the reagent itself — essential when reagents have inherent color or when lot-to-lot variation is significant. Without it, residual reagent color adds a systematic positive offset to every reading.

Repeat for every new reagent lot. The reagent blank correction is only valid for the specific lot of reagent used when the blank was entered. When a new lot is opened, always re-run the blank procedure before testing samples.
1

Select the applicable program

Navigate to and start the method you want to apply the correction to.

2

Press Options → Advanced Options → Reagent Blank

This opens the reagent blank entry mode.

3

Prepare a blank using deionized water instead of sample

Run the complete test procedure using deionized water in place of the actual sample — add the same reagent, observe the same reaction time, swirl as directed. Do not shortcut the procedure.

4

When the reading appears — press OK to store as the blank

The value displayed is the reagent’s contribution to absorbance at zero actual analyte concentration. Pressing OK stores it. The Reagent Blank icon now appears on the measurement screen — confirming the correction is active.

5

All subsequent readings are automatically corrected

The stored blank value is subtracted from every result for this program until a new blank is entered or the correction is cleared.

Standard Adjust — Step by Step

Standard Adjust modifies the stored factory calibration to align the instrument reading to a verified known standard. Use this when a standard solution at a known concentration gives a reading that differs from the expected value — indicating instrument drift, cell variation, or lot-specific reagent differences.

If using both Reagent Blank and Standard Adjust — Reagent Blank FIRST. Perform the Reagent Blank correction before applying Standard Adjust. Reversing the order produces an incorrect calibration offset.
1

Run a complete test on a known standard solution

Use a verified standard solution at a concentration near the top of the method range. Follow the complete test procedure including reagent addition and reaction time.

2

Press Options → Advanced Options → Standard Adjust

The display shows the current reading from Step 1.

3

Enter the known concentration of the standard — press OK

Use the navigation keys to enter the certified concentration. Press OK or Done to apply. The Standard Adjust icon appears on the display confirming the offset is active.

Standard Adjust vs SCA on the DR300: The DR900’s Standard Adjust is functionally equivalent to the DR300’s SCA (Standard Calibration Adjust) but applies across any method without the DR300’s reagent-specific standard range requirements. Use it any time a verified standard confirms the factory calibration has drifted.

Operator ID & Sample ID — Data Traceability

The Operator ID and Sample ID features are the DR900’s primary tool for regulatory data traceability — especially relevant for compliance monitoring programs where records must be attributable to a specific operator and sample location.

FeatureCapacityAppears InAccess Path
Operator IDUp to 10 IDsEvery data log entry; USB export CSVSETTINGS → Operator ID
Sample IDUp to 10 IDsEvery data log entry; USB export CSVSETTINGS → Sample ID
For multi-site or multi-operator programs: set up all Operator IDs and Sample IDs before field deployment. Operators select their ID from the pre-loaded list — no keyboard entry required in the field. The CSV export from the USB data log then includes fully attributed records for every measurement.

USB Functions — Data Export, Firmware, Programs

Connect the DR900 to a PC via the included USB mini cable. No software installation is required — the instrument appears as a standard USB drive. The USB Main Menu on the instrument display presents three options:

USB FunctionWhat It DoesFile Name / Notes
Data Log exportExports all 500 stored results to CSV files on the PCDR900_DataLog.csv (results) • DR900_EventLog.csv (events)
Firmware updateUpdates instrument firmware from a PC-downloaded fileFile must be named DR900.bin; select Update DR900 from USB menu
Program managementImport and export user programs between instruments or from PCEnables standardizing custom methods across a fleet of DR900s
Data export procedure: Select Data Log from the USB Main Menu on the instrument (not just the PC file manager). Use the instrument’s Disconnect USB option before physically unplugging the cable to prevent file corruption. Do not disconnect while a transfer is in progress.

Creating a User Program

User programs allow operators to store custom measurement methods not covered by the 90+ factory programs, or modified versions of existing methods. Maximum 10 programs across user programs and favorites combined.

1

Navigate to HOME → Options → All Programs → Options → User Programs

This opens the user program creation interface.

2

Select a wavelength: 420, 520, 560, or 610 nm

Choose the wavelength appropriate for the colorimetric chemistry of your method. Refer to the reagent manufacturer’s documentation for the correct measurement wavelength.

3

Enter calibration data (absorbance / concentration pairs)

Enter the calibration points that define the concentration curve for your method. More points across the range produce a more accurate calibration.

4

Name and save the program

Assign a name. The program is saved and appears in the Favorites / User Programs list. It can also be exported via USB and imported onto other DR900 instruments for fleet standardization.

All 48+ Parameters — Complete Method List

The DR900 comes pre-loaded with 90+ methods covering the following 48+ parameters. Wavelength is automatically selected by the instrument based on the method selected. All method numbers and ranges from the official DR900 Methods Reference (DR900_CMS_block.pdf, Hach).

Alkalinity, Total
0–500 mg/L CaCO₃
Method 10239
Ammonia Nitrogen
0.01–0.50 / 0.4–50 mg/L
Methods 8155, 10031
Bromine, Total
0.05–4.50 / 0.2–10.0 mg/L
Method 8016
Chlorine, Free
0.02–2.00 mg/L
Method 8021 (DPD)
Chlorine, Total
0.02–2.00 mg/L
Method 8167 (DPD)
Chlorine Dioxide
0.01–0.50 mg/L
Method 8138
Chromium, Hex
0.005–0.700 mg/L
Method 8023
Chromium, Total
0.01–0.70 mg/L
Method 8024
COD (Cr)
0–1500 mg/L
Methods 8000, 10211
Color, True/Apparent
0–500 PCU
Method 8025
Copper
0.04–5.00 mg/L
Method 8506
Cyanide
0.002–0.240 mg/L
Method 8027
Dissolved Oxygen
0–12 mg/L
Method 8166
Fluoride, SPADNS
0.02–2.00 mg/L
Method 8029
Hardness, Total
0–400 mg/L CaCO₃
Method 8213
Hydrogen Peroxide
0.02–1.00 mg/L
Method 8222
Iron, Total
0.02–3.00 mg/L
Method 8008 (FerroVer)
Iron, Dissolved
0.02–3.00 mg/L
Method 8008
Lead
0.001–0.100 mg/L
Method 8033
Manganese
0.006–0.700 mg/L
Method 8034
Mercury
0.001–0.100 mg/L
Method 8026 (Dithizone)
Molybdenum
0.01–40.0 mg/L
Method 8036
Monochloramine
0.04–4.50 mg/L
Method 10171
Nickel
0.01–0.70 mg/L
Method 8037
Nitrate, LR
0–0.50 mg/L NO₃−-N
Method 8039
Nitrate, HR
0–30.0 mg/L NO₃−-N
Method 8039
Nitrite
0.002–0.300 mg/L NO₂−-N
Method 8507
Ozone
0.01–1.00 mg/L
Method 8311 (Indigo)
pH, Colorimetric
6.5–8.5
Method 8025
Phosphate, Reactive
0.02–2.50 mg/L PO₄³−
Method 8048 (PhosVer 3)
Phosphorus, Total
0.06–3.50 mg/L PO₄³−
Method 8190
Potassium
0–8.00 mg/L
Method 8049
Residual Chlorine
0–10.0 mg/L
Method 10069 (HR)
Silica, LR
0.01–1.00 mg/L SiO₂
Method 8185
Silica, HR
1–100 mg/L SiO₂
Method 8186
Sulfate
0–70 mg/L
Method 8051
Sulfide
0.01–0.70 mg/L
Method 8131
Surfactants (Anionic)
0.05–0.500 mg/L
Method 8028
Thiocyanate
0.005–2.00 mg/L
Method 8000
Turbidity
0–1000 NTU
Method 8195
Zinc
0.01–3.00 mg/L
Method 8009

Full method list from DR900_CMS_block.pdf (Hach Methods Reference). Ranges and method numbers subject to change with firmware updates. Download the current methods reference from hach.com for the most current data.

Maintenance

Cleaning the Instrument Exterior

Clean the exterior with a moist cloth and mild soap solution, then wipe dry. The IP67 enclosure permits rinsing under running water. Do not use abrasive cleaners, solvents, or high-pressure water jets.

Cleaning Sample Cells

1

Rinse immediately after each use

Rinse cells with deionized water immediately after each test — do not allow reagents to dry in the cell. Dried reagent residue is the leading cause of background absorbance errors.

2

Mild soap and water for general cleaning

Use a mild laboratory detergent with warm water. Rinse thoroughly with deionized water. Dry with a lint-free cloth or air dry.

3

Method-specific cleaning requirements override general guidance

Certain methods require specific cleaning procedures (e.g., nitric acid rinse for trace metal methods; solvent rinse for organic methods). Always check the specific method procedure document first. Method-specific requirements take precedence.

4

Replace scratched or etched cells

Any visible scratch or etching in the optical path permanently affects readings. Replace cells that have been scratched by abrasive cleaners or physical damage.

Battery Replacement

Always replace all 4 AA batteries simultaneously. Do not mix new and used batteries or batteries from different brands. Do not use rechargeable batteries — the DR900 requires AA alkaline only. Replace when the battery icon indicates low power. Screw the battery compartment cover to 1–1.4 N·m torque when closing.

Troubleshooting

SymptomLikely CauseResolution
“+++” over-range displaySample concentration exceeds the upper limit of the selected method rangeDilute sample with reagent-demand-free water; retest; multiply result by dilution factor
“---” under-range displayValue below method’s lower detection limit; reagent not added; procedure errorVerify reagent was added; confirm procedure; check reagent expiry date; repeat test
High background reading (blank)Contaminated sample cell; reagent carryover from previous testClean or replace sample cell; triple-rinse with deionized water; run Reagent Blank correction
Inconsistent results between replicate testsReaction time not observed; swirling incomplete; cell not wiped; reagent lot variationStrictly follow method procedure timing; run Reagent Blank for new lot; verify standard with Standard Adjust
Standard Adjust reading outside expected rangeStandard concentration incorrect; standard expired; reagent lot changePrepare a fresh standard from a verified source; re-run Reagent Blank before Standard Adjust
Display will not power onDead or incorrectly installed batteriesVerify correct battery orientation; replace all 4 AA batteries simultaneously; check battery compartment screw
Adapter not seating correctlyCell adapter installed in wrong orientation or damagedRemove adapter, check orientation against Figure 3 in user manual, reseat
USB not recognized by PCCable not fully seated; instrument not powered on; firmware issueReconnect cable with instrument powered on; try a different USB port; check for firmware update

Replacement Parts & Accessories

ItemNotes
Sample cell, 16 mm glass test vialMost common — used with the cell adapter. Verify current item number at hach.com for your specific method.
Sample cell, 25 mm round glassCompatible with DPD chlorine, FerroVer iron, and other standard methods that use the 25 mm cell format.
Sample cell, 1 cm plastic (10 mL)Select methods only. Replace when scratched or cloudy.
Cell adapterRequired for 16-mm vials. Contact Hach if original is lost or damaged.
USB mini cableStandard USB mini-B. Any replacement USB mini-B cable is compatible.
Batteries, AA alkaline (4-pack)Standard AA alkaline. Do not use rechargeable.
Instrument case / protective coverContact Hach distributor or hach.com for current item numbers and regional availability.
Hach DR900 Multiparameter Portable Colorimeter — 9385100
90+ methods • 48+ parameters • IP67 • 500-result data log • USB export. Complete instrument without reagents.
Buy on Amazon →

Related DR300 Single-Parameter References

If you need a dedicated single-parameter instrument for one specific test, the DR300 family covers the most common parameters: