HomeReviews › Myron L Ultrameter II 6PIIFCE

Myron L Ultrameter II 6PIIFCE Review (2026)

Disclosure: This page contains affiliate links. If you purchase through our links, we may earn a commission at no extra cost to you. See our affiliate disclosure.

Reviewed by Lawrence Quarles, Alabama Grade IV Wastewater Treatment Operator
Myron L Ultrameter II 6PIIFCE multiparameter water quality meter — blue handheld instrument showing COND, RES, TDS, ORP, pH buttons and LCD display
CWL Verdict — The Professional Standard for Multiparameter Water Quality in the Field
The Ultrameter II 6PIIFCE is the benchmark handheld multiparameter meter for water treatment professionals who need reliable readings across seven parameters without carrying multiple instruments. The four-electrode conductivity cell, proprietary FCE reagent-free free chlorine measurement, 442™ Natural Water standard, factory-restorable calibration, and 100-point timestamped datalogger represent engineering decisions that have earned this instrument decades of trust in water treatment facilities, pool operations, and industrial applications.

Best for: Water treatment operators, pool/spa professionals, industrial process engineers, RO system technicians, municipal water compliance, and any application requiring simultaneous multiparameter monitoring with traceable data records.

Made in Carlsbad, California. NIST-traceable calibration. IP67/NEMA 6. ±1% conductivity accuracy.
Parameters
7: COND, RES,
TDS, pH, ORP,
FCE™, Temp
COND Accuracy
±1% of reading
pH Range
0–14 pH
±0.01 pH res.
IP Rating
IP67 / NEMA 6
1 m, 30 min
Memory
100 points
RTC timestamped
Cell Type
Four-electrode
(Kelvin)
Standards
KCl · NaCl
442™ · User
Origin
Made in USA
Carlsbad, CA
Myron L Ultrameter II 6PIIFCE — ASIN B015JP23R8
7 parameters • ±1% accuracy • IP67/NEMA 6 • FCE™ free chlorine • 100-point datalogger • Made in USA
View on Amazon →

7 Parameters Explained

COND key
Electrical Conductivity
Measures a solution's ability to carry electrical current — a direct proxy for total ionic concentration. Units: µS (microsiemens) or mS. Automatically temperature-compensated to 25°C. ±1% accuracy.
RES key
Resistivity
The reciprocal of conductivity — used for high-purity water measurement in pharmaceutical, semiconductor, and boiler applications. Range: 10 KΩ–30 MΩ·cm. Flowing sample method required for accurate high-purity readings.
TDS key
Total Dissolved Solids
Derived from temperature-compensated conductivity via a solution-specific conversion ratio. Units: ppm or ppt. Accuracy depends on correct solution type selection — a 40% TDS error is possible between NaCl and 442 mode at the same conductivity reading.
pH key
pH
Glass membrane + Ag/AgCl reference junction in a single replaceable sensor assembly. Range 0–14 pH, 0.01 resolution. Three-point calibration: pH 7 zero, then pH 4 (acid) and/or pH 10 (base). Always calibrate pH 7 first.
ORP/Fr Chl key
ORP (Redox)
Platinum electrode measures oxidizing or reducing tendency of the solution in millivolts. Positive = oxidizing (chlorinated water). Negative = reducing. Zero calibrated electronically during pH 7 calibration — no separate ORP standard solution required.
ORP/Fr Chl key
FCE™ Free Chlorine
Proprietary reagent-free free chlorine measurement: ORP + pH → ppm via laboratory-derived conversion table. Non-destructive. Reports actual sanitizing capability. No DPD reagents, no color comparison, no sample alteration. Pool, spa, and drinking water disinfection verification.
Auto-displayed
Temperature
Always shown on secondary display during all measurements. Used automatically by the microprocessor for temperature compensation of all conductivity, resistivity, and TDS readings. Reference temperature: 25°C.

Four-Electrode Conductivity Cell

Most entry-level conductivity meters use a two-electrode design, where the same electrodes that drive current through the sample also measure the voltage response. This creates a systematic error: the electrode-solution interface has its own impedance (contact resistance), which is indistinguishable from the sample's resistance in a two-electrode measurement. The result is that fouled, aged, or partially blocked electrodes read incorrectly — and the operator may not know it.

The Ultrameter II uses a four-electrode (Kelvin) configuration. Two outer electrodes drive an AC current through the sample; two separate inner electrodes measure the resulting voltage. Because the inner measurement electrodes carry essentially no current, the electrode-solution contact impedance has negligible effect on the reading. The four-wire arrangement delivers ±1% accuracy regardless of electrode surface condition, fill level variations, or minor fouling — a design choice that directly explains why this instrument maintains calibration stability longer than two-electrode alternatives.

FCE™ Free Chlorine — How It Works

Standard DPD colorimetric testing adds a reagent to the sample that reacts with chlorine species to produce a pink color whose intensity is compared to a color chart or read by a photometer. The process chemically alters the sample, cannot be repeated on the same aliquot, requires reagent handling and disposal, and — critically — cannot distinguish between the sanitizing effectiveness of chlorine at different pH values.

The FCE approach exploits a physical chemistry relationship: in chlorinated water, ORP is a function of both the free available chlorine (FAC) concentration and the pH. Because FAC exists in two forms (hypochlorous acid, HOCl, and hypochlorite ion, OCl⁻) whose ratio is governed by pH, and because HOCl is approximately 80–100 times more effective as a disinfectant than OCl⁻, a measurement that captures both ORP and pH captures actual sanitizing power — not just total chlorine concentration.

Why FCE is more informative than DPD for pool and spa operators: DPD reports free chlorine concentration in ppm regardless of pH. At pH 7.0, most FAC is in the effective HOCl form. At pH 8.0, most FAC is in the less effective OCl⁻ form. A DPD reading of 3 ppm at pH 7.0 and 3 ppm at pH 8.0 appear identical — but the sanitizing power at pH 7.0 is several times greater. FCE converts ORP + pH simultaneously into an effective chlorine ppm that reflects this reality. Higher FCE ppm = more effective disinfection, accounting for pH.

The FCE conversion table was derived from controlled laboratory experiments with precisely calibrated chlorine levels and excluded interferants. The conversion applies to free available chlorine (hypochlorous acid and hypochlorite ion) specifically and is not valid for combined chlorine (chloramines) or other oxidants.

The 442™ Natural Water Standard

Conductivity measurements are calibrated against reference solutions of known conductivity. The choice of reference solution matters because different ionic species contribute differently to conductivity per unit mass — meaning a 1,000 µS reading in KCl solution represents a different TDS than the same 1,000 µS in a natural freshwater sample.

Myron L's 442™ Natural Water standard — 40% sodium sulfate, 40% sodium bicarbonate, 20% sodium chloride — was formulated to replicate the ionic balance of typical fresh natural water. When the Ultrameter II is calibrated with the 442 standard and the 442 solution type is selected for TDS conversion, the resulting ppm reading correlates closely with gravimetric TDS (evaporated weight measurement) used in regulatory reporting. For water treatment, drinking water, and environmental monitoring applications, this is a meaningful accuracy advantage over KCl-calibrated instruments.

Full Specifications

ParameterRangeResolutionAccuracy
Conductivity (COND)0–199.9 µS / 200–1999 µS / 2.00–19.99 mS / 20.0–199.9 mS0.1 µS / 1 µS / 0.01 mS / 0.1 mS±1% of reading
Resistivity (RES)10 KΩ–30 MΩ·cmAuto-ranging±1% of reading
TDSDerived from COND (solution-type dependent)Depends on range±1% of reading (conductivity basis)
pH0–14 pH0.01 pH±0.01 pH (buffer calibrated)
ORP±1999 mV1 mVPlatinum electrode
FCE™ Free Chlorine0.01–5.00 ppm Cl₂0.01 ppmDerived from ORP + pH
Temperature0–100°C0.1°CCompensated reference for COND/TDS/RES
Source: Myron L Company Ultrameter II Operation Manual (UMIIFCEOM 18FE16, 2016). Conductivity accuracy ±1% of reading — not ±% of full scale. Factory calibration NIST-traceable.
Physical / EnvironmentalSpecification
IP ratingIP67 / NEMA 6 — dust-tight, submersible to 1 meter for 30 minutes
Battery9V alkaline
Auto-off15 seconds default (adjustable to 75 sec); 60 sec in calibration mode
Memory100 readings — EEPROM non-volatile; real-time clock date/time stamp on each record
Calibration storageNon-volatile; factory calibration (FAC) permanently stored and restorable
Solution typesKCl, NaCl, 442™ Natural Water, User (custom Tempco and TDS ratio)
EMC complianceCE certified; IEC 61000-4-2, IEC 61000-4-3, CISPR 11, EN61326-1
OriginMade in USA — Carlsbad, California

Solution Selection and User Mode

The solution type selection controls two linked calculations simultaneously: the temperature compensation model applied to the raw conductivity reading, and the conversion ratio used to derive TDS from that compensated conductivity. The Ultrameter II maintains separate calibration data for each solution type — changing the selection does not affect calibrations performed for other types.

Solution TypeBest Used ForTDS Basis
KClLaboratory standard reference; conductivity calibration verificationKCl ionic weight
NaClSeawater, brine, desalination applications; cooling tower blowdown; food-grade salt solutionsNaCl ionic weight
442™ Natural WaterDrinking water, freshwater, municipal water treatment, RO permeate, environmental monitoring — correlates with gravimetric TDS40% Na₂SO₄ / 40% NaHCO₃ / 20% NaCl
UserAny solution with known TDS but unique ionic composition — hydroponic nutrient solutions, specialty chemical processes, industrial process streamsCustom ratio (operator-programmed)
Factory defaults: KCl for conductivity, NaCl for resistivity, 442 for TDS. The 40% TDS error example (NaCl vs. 442 at the same µS reading) illustrates why correct solution selection is not optional for accurate TDS reporting.

Setting a Custom User Mode TDS Ratio

To program a custom conductivity-to-TDS conversion ratio: measure the solution at 25°C in COND mode and record the µS reading. Divide the solution's known TDS (ppm) by the measured µS. For example: 75 ppm TDS at 100 µS = ratio of 0.75. Enter this ratio (0.20–7.99) in User mode via the CAL key sequence. The custom Tempco (%/°C, range 0.00–9.99) should also be entered for the specific solution. Setting Tempco to 0.00 disables temperature compensation.

Operating Procedures — All Parameters

Conductivity and TDS

1

Verify solution type

Check the left side of the LCD for the current solution type (KCl, NaCl, 442, or User). To change: press COND or TDS, hold CAL for 3 seconds until "SEL" appears, use ▲▼ to select, press CAL to confirm.

2

Rinse the cell cup three times

Rinse the conductivity cell cup three times with the test sample — not distilled water, not rinse water. This conditions the temperature compensation network and removes contamination from previous measurements. Three rinses, not one.

3

Fill and measure

Refill the cell cup with the test sample. Verify no air bubbles are clinging to the cell walls or electrodes — bubbles cause falsely low readings. Press COND or TDS. Allow reading to stabilize.

4

Store and rinse

Press ▲ MS to store the reading with timestamp. After measurement, rinse the cell with clean water to prevent crystal formation on the electrodes.

Resistivity (High-Purity Water)

Pure water absorbs CO₂ from the atmosphere within seconds of contact, driving resistivity readings artificially downward. Static sample measurement is not reliable for high-purity water. Follow the flowing sample method:

  1. Confirm the pH protective cap is secured on the pH/ORP sensor well.
  2. Hold the instrument at a 30° angle with the cell cup sloping downward.
  3. Allow sample to flow continuously into and out of the cell without aeration.
  4. Press RES and observe the stabilizing reading — use the most stable value observed.

pH Measurement

1

Remove the pH/ORP protective cap

Rotate while grasping and pulling upward to remove. The cap contains pH sensor storage solution — avoid spilling.

2

Rinse both sensor well AND cell cup — three times

Rinse both the pH/ORP sensor well and the conductivity cell cup three times with the test sample. Shake out each rinse completely before the next. Both must be rinsed — contamination in either chamber affects the reference junction.

3

Fill and measure

Refill both sensor well and cell cup with test sample. Press pH and allow reading to stabilize.

!

CRITICAL — Store the pH sensor after every use

After use, fill the pH/ORP sensor well with Myron L pH Sensor Storage Solution and replace the protective cap securely. Storing the sensor dry causes the reference junction KCl to crystallize, blocking the junction and producing drift or unresponsive pH readings. This is the most common cause of pH sensor failure on this instrument.

ORP Measurement

Confirm ORP mode is active (not FCE mode): press ORP/Fr Chl, hold CAL 3 seconds, verify display shows "mV" not "ppm". Use ▲▼ to toggle if needed, then press any parameter key to save. Procedure: remove cap → rinse sensor well and cell cup three times → fill both → press ORP/Fr Chl → read stable mV value → press MS to store → fill sensor well with storage solution, replace cap.

FCE Measurement — Flow and Immersion Methods

Flow Method — Recommended for Discrete Samples

Set auto-off to 75 seconds before starting FCE measurements. FCE stabilization typically takes 1–2 minutes. The default 15-second auto-off will shut the instrument down mid-measurement. Navigate: press ▼ MR → scroll to Auto oFF → press CAL → use ▲▼ to select 75 sec → press CAL.
1

Set FCE mode

Press ORP/Fr Chl → hold CAL 3 seconds → use ▲▼ to select "Chl" ppm (not "ORP" mV) → press any parameter key to save.

2

Empty sensor well of storage solution

Remove cap and empty the storage solution from the pH/ORP sensor well before flushing. Storage solution residue dilutes the sample and produces falsely low FCE readings.

3

Flush for 10+ seconds

Hold at 30° angle (cell cup sloping down). Flush sensor well and cell cup with a steady sample stream for at least 10 seconds. Allow sample to flow continuously — do not aerate.

4

Press ORP/Fr Chl and wait for stabilization

Display alternates between predicted ORP (mV) and FCE ppm. Wait until both readings are unchanged for 5 consecutive readings — typically 1–2 minutes. If it takes over 1 minute, press ORP/Fr Chl again to reset the auto-off timer.

5

Record and repeat

Press MS to store the stable FCE ppm reading. Repeat the full procedure twice more. Average all three readings for the reported value.

Immersion Method — Pools, Spas, and Large Bodies

  1. Hold the instrument beneath the water surface — surface chemistry differs from bulk water.
  2. Swirl the instrument for at least 10 seconds to flush cell cup and sensor well.
  3. Continue holding under the surface throughout measurement.
  4. Press ORP/Fr Chl. Wait for mV and ppm to stabilize (5 consecutive unchanged readings; 1–2 minutes typical).
  5. Press MS to store. Repeat twice and average the three readings.

Calibration Protocols

The Ultrameter II is factory-calibrated with NIST-traceable Myron L standard solutions before shipment. The microprocessor architecture eliminates mechanical drift, extending calibration intervals significantly versus older designs. Factory calibration values are stored permanently in non-volatile EEPROM and can be restored at any time.

Conductivity / TDS Calibration

  1. Rinse the conductivity cell cup three times with the appropriate standard solution (KCl, NaCl, or 442).
  2. Refill the cell cup with the same standard.
  3. Press COND or TDS to begin measuring.
  4. Press CAL — the CAL annunciator appears on the LCD.
  5. Use ▲▼ keys to adjust the displayed value to match the standard solution's labeled value.
  6. Press CAL to accept and complete calibration for this solution type.
  7. To calibrate additional solution types, change the solution type and repeat from step 1.

pH Calibration — Three-Point Procedure

Always calibrate pH 7 (zero point) first. The pH 7 calibration sets the ORP zero reference simultaneously — it cannot be performed after gain calibration. Calibrating pH 4 or pH 10 before pH 7 produces an incorrect calibration.
  1. Rinse sensor well and cell cup three times with pH 7 buffer. Fill with pH 7 buffer.
  2. Press pH. When stable, press CAL. Adjust to 7.00. Press CAL to save. This also calibrates the ORP zero point electronically.
  3. Rinse three times with pH 4 buffer (acid gain). Fill with pH 4 buffer. Press pH. When stable, press CAL. Adjust to 4.00. Press CAL.
  4. Optionally: rinse three times with pH 10 buffer (base gain). Fill with pH 10. Press pH. When stable, press CAL. Adjust to 10.00. Press CAL.

100-Point Datalogger and Real-Time Clock

Every stored record includes: measured value and parameter, measurement units, solution type in use, temperature at time of measurement, and a real-time clock time and date stamp. Memory is non-volatile — stored data and all calibration settings survive battery replacement provided the battery is removed for fewer than 3 minutes.

OperationProcedure
Save readingWhile desired measurement is displayed, press ▲ MS. MEMORY annunciator flashes; temperature display momentarily shows the memory location number (1–100).
Recall readingsPress ▼ MR in any measurement mode. Scroll through records with ▲▼. Press CAL while viewing a record to display its full time and date stamp.
Delete individual recordWhile the record is displayed in Recall mode, press and hold CAL/MCLR.
Delete all recordsScroll past location 100 to "CLr ALL" → press CAL to confirm. All 100 records erased.
Configure real-time clockAccess via Memory Recall scroll menu. Time in 24-hour format. Date format: US (MM/DD/YY) or International (DD/MM/YY) — selectable independently.
Memory fills locations 1–100 sequentially. Once full, new records overwrite the lowest-numbered empty or cleared location. Non-volatile EEPROM — battery changes do not erase stored data if battery is replaced within 3 minutes.

Cell Check and Factory Calibration Restore

Cell Check Function

Cell Check electronically verifies the conductivity cell's cleanliness without requiring standard solution. It is the first step in any conductivity troubleshooting sequence.

Access: press ▼ MR → scroll past FAC SEL → "CELL ch" appears → press CAL to run the test.
Result "Good": cell is clean — proceed with measurement or calibration.
Result "CELL cLn": cell requires cleaning before accurate measurements are possible. Clean with isopropyl alcohol or foaming non-abrasive household cleaner. Do not scrub cell interior. Rinse, recheck.

Factory Calibration Restore (FAC)

If field calibration is suspected to be incorrect and no standard solution is available, the original factory calibration can be restored from permanent EEPROM storage.

Access: press ▼ MR → scroll past CLr ALL → "FAC SEL" appears → press CAL to execute.

FAC restore is a recovery step, not a substitute for calibration. FAC reload restores the instrument's internal electronic calibration — not a solution-referenced calibration against a physical standard. Use FAC restore when field calibration is clearly wrong and no standard is available. As soon as standard solutions are accessible, perform proper solution-referenced calibration to verify and correct.

Maintenance and Troubleshooting

Conductivity Cell Cleaning

Rinse the cell with clean water after every use. Discoloration of the stainless steel electrodes from chemically active samples does not impair accuracy — do not clean purely for appearance. When visible oil, scale, or fouling films are present, clean with isopropyl alcohol or a foaming non-abrasive cleaner (not abrasive scrubbers). Run Cell Check after cleaning to verify the cell is restored to "Good" status before taking measurements.

pH/ORP Sensor Rejuvenation

A sensor stored without the cap, or exposed to extreme heat, may develop a blocked reference junction from crystallized KCl. Two rejuvenation methods:

Battery Replacement

Dry the instrument completely before opening the case. Water inside the case will short the electronics. Remove all four bottom screws, carefully open the case, replace the 9V alkaline battery via the circuit board connector, ensure the sealing gasket is properly seated in its groove before closing, and reinstall all four screws tightened evenly to maintain the IP67 waterproof seal.

Applications by Sector

ApplicationPrimary ParametersKey Capability
Municipal drinking waterTDS (442 mode), FCE ppm, pHFCE verifies FAC residuals without DPD reagent handling. 442 mode TDS correlates with regulatory evaporated-weight measurements.
Pool and spaFCE ppm, ORP, pHImmersion method provides bulk water readings. Three simultaneous parameters — FAC, ORP, and pH — in one measurement cycle. No reagent disposal.
RO and water purificationTDS (feed and permeate), ResistivityMemory logging with timestamps enables membrane performance trending. User mode for custom feed water ionic profile.
Industrial boiler / high-purityResistivity (10 KΩ–30 MΩ·cm)Flowing sample method for high-purity water prevents CO₂ interference. Resistivity range covers semiconductor and pharmaceutical process water.
Cooling towersCOND, TDS (NaCl mode), ORPNaCl mode appropriate for cooling tower blowdown chemistry. ORP monitors biocide effectiveness.
Agriculture / hydroponicsEC (µS/mS), TDS (User mode)User mode custom TDS ratio matches specific nutrient solution chemistry. EC as nutrient concentration proxy.
Myron L Ultrameter II 6PIIFCE — ASIN B015JP23R8
7 parameters • ±1% accuracy • IP67/NEMA 6 • FCE™ free chlorine • 100-point RTC datalogger • Made in USA • NIST-traceable
View on Amazon →

FAQ

What parameters does the Ultrameter II 6PIIFCE measure?

Seven: conductivity, resistivity, TDS, pH, ORP, FCE™ free chlorine (ppm), and temperature. All displayed simultaneously — temperature always appears on the secondary display. The Model 4P variant measures conductivity, resistivity, TDS, and temperature only (no pH, ORP, or FCE).

What is FCE and how does it differ from DPD?

FCE (Free Chlorine Equivalent) is Myron L's reagent-free free chlorine measurement. The instrument simultaneously measures ORP and pH, then applies a proprietary laboratory-derived conversion to calculate effective free chlorine ppm. Unlike DPD, FCE doesn't alter the sample, doesn't require reagents, and accounts for pH when calculating sanitizing effectiveness. A full explanation is in the FCE section above.

What is the 442 Natural Water standard?

40% sodium sulfate, 40% sodium bicarbonate, 20% sodium chloride — a blend formulated by Myron L to replicate the ionic profile of fresh natural water. Using 442 mode produces TDS readings that correlate with gravimetric TDS measurements used in regulatory reporting. More accurate than NaCl or KCl mode for drinking water and freshwater applications.

How do I run the Cell Check function?

Press ▼ MR → scroll past FAC SEL → "CELL ch" appears → press CAL. "Good" = cell is clean. "CELL cLn" = clean the cell before measuring. No standard solution required.

Can the factory calibration be restored?

Yes — permanently stored FAC values in EEPROM can be restored any time: press ▼ MR → scroll past CLr ALL → "FAC SEL" → press CAL. This is a recovery option when field calibration is suspected to be wrong. It is not a substitute for calibration against NIST-traceable standard solutions.

How does User mode work?

User mode allows programming a custom temperature compensation coefficient (Tempco, 0.00–9.99 %/°C) and custom conductivity-to-TDS conversion ratio (0.20–7.99) for solutions that don't match KCl, NaCl, or 442. To determine the ratio: measure the solution's conductivity in µS at 25°C, then divide the known TDS (ppm) by that µS value. See the solution selection section above for the full procedure.

Related Reviews