HomeGuides › Measuring TDS in Water

Measuring TDS in Water: What Your Meter Actually Measures

Reviewed by Lawrence Quarles, Alabama Grade IV Wastewater Treatment Operator · Updated July 2026

Start with the fact that reframes everything else on this page: no handheld TDS meter measures total dissolved solids. Every one of them — the $8 pen and the $1,089 Ultrameter alike — measures electrical conductivity and multiplies by a conversion factor to produce the ppm number on the display. That's a legitimate and widely accepted method. But the conversion factor isn't a constant, most inexpensive meters don't disclose which one they use, and the difference between the common factors is large enough to change decisions. This guide covers what the measurement actually is, how much error the conversion introduces, how to calibrate and run a field protocol that produces defensible numbers, and how to interpret the result.

What TDS is, and the reference method

Total dissolved solids is the combined concentration of everything dissolved in a water sample — minerals, metals, salts, and dissolved organic compounds. The operational definition used across the extension literature is material smaller than 2 microns that cannot be removed by conventional filtration. Natural contributors include calcium, magnesium, sodium, potassium, carbonates, bicarbonates, chlorides, sulfates, nitrates, and silica, leached from soil and rock. Anthropogenic contributors include road salt, agricultural runoff, treatment chemicals, pipe corrosion products, and process water carryover.

The reference method is gravimetric and conceptually simple: take a measured sample, evaporate the water to dryness, and weigh what's left. Standard Methods specifies drying at 180 °C, and results are reported in mg/L. It's accurate, it's the number every other method is trying to approximate, and it's completely impractical in the field — it needs a laboratory balance, a drying oven, and hours per sample.

That impracticality is the entire reason conductivity meters exist. Everything that follows is about how good the substitute is.

Clear water tells you nothing. Dissolved solids are, by definition, dissolved — they don't scatter light and they don't settle out. SDSU Extension found in field testing that more samples that looked clear tested poorly than samples that looked murky. Visual assessment of a water source is not a screening method, and neither are the folk heuristics that circulate alongside it: waterfowl on a pond and spring-fed sources are both commonly cited as signs of good water, and neither holds up.

What the meter actually measures

Dissolved minerals dissociate into ions, and ions carry charge. That makes conductivity a workable proxy for dissolved ion concentration: the more dissolved mineral content, the more readily water conducts. Distilled water conducts poorly precisely because there's almost nothing dissolved in it to carry current.

An EC meter applies a voltage across two electrodes at a known spacing and geometry — the cell constant — and measures the resulting current. That gives conductivity, in microsiemens or millisiemens per centimeter. A "TDS meter" is the same instrument with one extra step in firmware: it multiplies the conductivity reading by a conversion factor and displays the product as ppm.

What the display implies

"This water contains 340 ppm of total dissolved solids." Presented as a measurement, in the same units a laboratory would report.

What actually happened

The meter measured roughly 520 µS/cm of conductivity and multiplied by a factor — probably 0.65, possibly 0.50, possibly 0.70, and on most inexpensive meters the documentation doesn't say which. The ppm figure is a derived estimate carrying every bit of the conversion's uncertainty on top of the instrument's own.

This is not a criticism of the method. Conductivity-derived TDS is genuinely useful, it's what extension programs recommend for field screening, and for trend monitoring it's entirely sufficient. The problem is specifically the undisclosed factor, which is where a meaningful and avoidable error enters.

The EC-to-TDS conversion table

The conversion factor is not a constant. It increases with conductivity, because the relationship between ion concentration and conductivity in natural water is non-linear — as ionic strength rises, ion–ion interactions reduce the conductivity contributed per unit of dissolved mass. The table below is from NDSU Extension publication WQ1923, corroborated by SDSU Extension, and is the standard reference for natural waters.

EC (µS/cm or µmhos/cm)EC (mS/cm)Approx. TDS (mg/L)Conversion factor
1000.1500.50
5000.53000.60
1,0001.06500.65
1,5001.51,0500.70
2,0002.01,4500.72
2,5002.51,8500.74
3,0003.02,2500.75
3,5003.52,6500.76
4,0004.03,0500.77
4,5004.53,5000.78
5,0005.03,9500.79
6,0006.04,7400.79
6,5006.55,1350.79
7,0007.05,6000.80
7,5007.56,0750.81
10,00010.08,2000.82
Source: NDSU Extension WQ1923 (Prasad & Meehan, revised January 2025); corroborated by SDSU Extension (Salverson, 2023). Factors apply to natural waters. The 3,500 µS/cm row is highlighted because it is the calibration point used in the field NaCl method below.

The factor spans 0.50 to 0.82 across the range — a 64% spread from bottom to top. Any instrument that applies one number to that whole curve is making an approximation whose size depends on where in the range you're working.

How wrong a fixed factor gets

Inexpensive pen meters typically hard-code a single conversion factor. The two conventions in common use are the NaCl scale at roughly 0.50 and the 442 scale at roughly 0.70, named for a reference mixture of 40% sodium sulfate, 40% sodium bicarbonate, and 20% sodium chloride. Some meters use 0.65. Many don't document the choice anywhere.

It gets worse when you cross industries, because different fields standardized on different numbers. Agricultural extension guidance for irrigation water commonly uses EC × 640 — a factor of 0.64 — for water below 5 dS/m. Hydroponic nutrient charts are usually built on the 500 or 700 scale. Well-water guidance uses the sliding factor in the table above. The consequence is that a single water sample can be correctly reported as four different ppm values by four sources that are each following their own accepted standard:

ConventionFactorWhere it is used1,000 µS/cm reads as
NaCl scale0.50Many inexpensive pen meters; hydroponics "500 scale"500 ppm
Agricultural / irrigation0.64Extension guidance for irrigation and greenhouse water640 ppm
NDSU sliding factor0.65Natural waters at this conductivity — the reference table above650 ppm
442 scale0.70Myron L and similar instruments; hydroponics "700 scale"700 ppm
Four accepted conventions, one water sample, a 40% spread between the lowest and highest reported value. None of these is wrong — they are answering slightly different questions about what "dissolved solids" means. This is the strongest practical argument for recording EC and stating the factor whenever you report ppm.

Taking the NDSU table as the reference and applying each fixed factor across the range gives the error directly:

EC (µS/cm)NDSU TDS (mg/L)Correct factorMeter fixed at 0.50 readsErrorMeter fixed at 0.70 readsError
100500.50500%70+40%
5003000.60250−17%350+17%
1,0006500.65500−23%700+8%
2,0001,4500.721,000−31%1,400−3%
3,5002,6500.761,750−34%2,450−8%
5,0003,9500.792,500−37%3,500−11%
10,0008,2000.825,000−39%7,000−15%
Calculated from the NDSU WQ1923 reference table. Each fixed-factor meter is exactly correct at one point on the curve and diverges in both directions from there. This is conversion error only — instrument accuracy, calibration drift, and temperature compensation error are additional and independent.

The practical shape of this: a 0.50-scale meter is accurate on very low-TDS water and reads progressively low as TDS rises, understating by nearly 40% on brackish water. A 0.70-scale meter is accurate in the 1,000–3,500 µS/cm band — which covers a lot of ordinary well and municipal water — and overstates badly at the clean end, which is exactly where RO permeate sits.

Where this actually bites. Two failure modes matter operationally. First, comparing readings from different meters: two instruments can both be working correctly and disagree by 40% because they use different scales, which makes cross-referencing a supplier's number against your own meaningless unless both scales are known. Second, sitting near a decision threshold: NDSU's field trigger for submitting a laboratory sample is TDS ≥ 4,500 ppm. Water at 6,000 µS/cm is about 4,740 mg/L by the reference table — over the threshold. A 0.50-scale meter displays 3,000 ppm for that same water and you don't send the sample.
The fix is straightforward. Work in EC, not ppm. Conductivity is the actual measurement — it's what the instrument does, it's reproducible across meters, and every published agricultural and industrial reference table is available in EC units. Convert to TDS only when you need to compare against a ppm-denominated standard, and do it deliberately using the table above rather than letting undisclosed firmware do it for you. If your meter allows the conversion factor to be set, set it to match your working range; if it doesn't, record the EC value.

What conductivity can't see

There's a second and more fundamental gap between conductivity-derived TDS and gravimetric TDS, separate from the conversion factor: conductivity only detects things that carry charge.

Dissolved solids that are non-ionic contribute mass to a gravimetric TDS result but contribute little or nothing to conductivity. Sugars, alcohols, many dissolved organic compounds, and undissociated silica all fall into this category. Water carrying a meaningful non-ionic dissolved load can return a conductivity reading that implies far less dissolved material than the evaporation method would find.

For most natural waters this hardly matters — the dissolved load is overwhelmingly ionic and the proxy holds well. It matters in specific cases:

EC units, decoded

Conductivity units are a persistent source of confusion, largely because two naming systems coexist — the older mho (ohm reversed) and the SI siemens — and they're numerically identical.

1 mmho/cm = 1 mS/cm = 1,000 µmhos/cm = 1,000 µS/cm

"Micro" is one-millionth; "milli" is one-thousandth. Older extension and agricultural literature tends to use µmhos/cm; modern instruments display µS/cm or mS/cm. When you're comparing a reading against a published table, confirm which unit the table uses — the difference between mS/cm and µS/cm is a factor of 1,000, and it's the single most common error in reading these tables.

TDS units are simpler: mg/L and ppm are equivalent for dilute aqueous solutions, and ppt means parts per thousand, so 1 ppt = 1,000 ppm.

Meter selection — the specs that matter

SDSU Extension's guidance on meter purchasing is blunt and correct: don't buy the cheapest one. But "buy a better one" needs specifics, because the marketing language on inexpensive meters is designed to sound like specification without being any.

"Lab grade accuracy"
Not a specification. No standard defines it, no body certifies it, and it appears on meters costing under $10. If a listing uses this phrase without stating accuracy as a percentage of reading, it has told you nothing measurable.
"Professional" / "high precision"
Same category. Precision and accuracy are different properties, and neither is established by an adjective. Look for a number with a unit attached.
Accuracy, as % of reading
The real spec. Commercial field instruments typically state ±1–2% of reading; laboratory-grade instruments ±0.5% or better. Note that "% of reading" and "% of full scale" are very different claims — full-scale specs look better and mean less.
Temperature compensation
Conductivity varies roughly 2% per °C. ATC is essential, but ask what compensation coefficient it uses and whether it's adjustable — a fixed coefficient tuned for one water type introduces error in another.
Adjustable conversion factor
Whether you can set the TDS factor, or at minimum read raw EC. This is the difference between a meter that reports a measurement and one that reports an undisclosed calculation.
Range and resolution
SDSU recommends field meters reading to at least 20.00 mS/cm for agricultural work. Resolution of 1 µS/cm is adequate for most commercial screening; 0.1 µS/cm for precision work.
Cell constant
Determines the usable range and the linearity across it. A K=1.0 cell suits mid-range water; K=0.1 suits high-purity; K=10 suits brine. Rarely stated on consumer meters, always stated on instruments.
Calibration certificate
Traceability to recognized standards. This is what makes a reading defensible in a compliance or dispute context, and it's most of what separates instrument pricing tiers.

Established instrument manufacturers in this space include Myron L, Hanna Instruments, YSI, Apera, Milwaukee, and Extech. That doesn't make every product from them right for every job, but it does mean published specifications, documented cell constants, available calibration standards, and service support.

Matching instrument to task, honestly. A $10 pen meter is a reasonable tool for trend monitoring — watching whether RO permeate is drifting upward week over week, confirming a softener is still doing something, checking a nutrient reservoir against yesterday. Trends are internally consistent even when the absolute number carries conversion error, because the error is systematic. The pen becomes the wrong tool the moment the number has to leave the building: regulatory reporting, a supplier dispute, a warranty claim, or anything a third party will act on. That threshold — not water type, not budget — is what determines which tier you actually need.

Calibration, including the field method

An uncalibrated meter produces numbers that look every bit as authoritative as calibrated ones. Conductivity cells age, electrode surfaces foul, and cell geometry shifts with physical shock — all of which move the reading without any indication on the display. NDSU Extension is explicit that meter accuracy should be verified against a calibration solution before sampling.

When to calibrate

Commercial calibration standards

The reliable approach is a purchased standard of certified conductivity. Common values are 84, 1,413, and 12,880 µS/cm — choose the one nearest your working range, since calibrating at 84 µS/cm and then measuring brackish water verifies almost nothing about performance at the top of the range. For drinking-water and food/beverage work, 1,413 µS/cm is the usual choice because it sits near the EPA secondary standard range.

The NDSU field method using table salt

When a commercial standard isn't available, NDSU WQ1923 documents a validated field procedure producing a solution near 2,600 mg/L TDS. That target wasn't arbitrary — it's approximately the level at which livestock begin showing adverse effects, and NDSU notes the conversion factor changes little above 2,700 mg/L, so a meter verified there reads reliably across the whole high range.

StepActionDetail
MaterialsTable salt, ¼-tsp measure, measuring cup, mixing container, distilled water, EC meterUse plain table salt rather than sea or kosher salt for repeatability. Use distilled water — tap water carries its own dissolved load and invalidates the standard.
Measure saltOne level ¼ teaspoon, not packedApproximately 1.25 mL by volume, roughly 1.7 g.
Measure waterExactly 4 cups (0.94 L) distilled waterMeasure on a level surface. Volume error here propagates directly into the standard's concentration.
Mix and restStir until fully dissolved, then rest 5 minutesSwirl gently rather than agitating. Undissolved salt produces a non-homogeneous solution and a low reading.
Set factorSet the meter's TDS conversion factor to 0.76Corresponds to the 3,500 µS/cm row of the NDSU table, matching this solution's strength.
Accept ifEC reads 3,200–3,600 µS/cm, or TDS reads 2,400–2,700 mg/LThe range reflects variability in table salt — iodine and anti-caking agents affect conductivity slightly. Outside this range, recalibrate per manufacturer instructions or service the meter.
DiscardMake fresh each timeDo not store and reuse. Evaporation and container interaction shift the concentration.
Source: NDSU Extension WQ1923 (Prasad & Meehan, revised January 2025). This is a verification and field-calibration method; it does not substitute for a traceable certified standard where documentation is required.

The same discipline applies to any field instrument — the pH meter calibration guide covers the equivalent protocol for electrometric pH, including slope acceptance criteria and compliance documentation.

Field testing protocol

This protocol combines NDSU WQ1923 and SDSU Extension guidance with standard industrial practice. It applies to wells, ponds, streams, process water, cooling tower basins, and boiler feedwater alike.

  1. Calibrate first, and log it. Verify against a standard before collecting any data, and record the result. A calibration record is what makes the day's readings defensible later.
  2. Fix and document sampling locations. For any monitoring program, establish permanent sampling points with GPS coordinates or unambiguous physical descriptions. Trend data from inconsistent locations isn't trend data.
  3. Rinse the container three times with the water being sampled before taking the final sample. Residue from a previous sample is a real and common source of error.
  4. Collect representatively. Surface water: scoop through the water column rather than skimming the surface. Wells: sample from active flow, not water standing in the pipe. Process water: mid-stream, never from a dead leg.
  5. Let the reading stabilize — typically 15–30 seconds while the probe equilibrates to sample temperature. Reading before stabilization is reading the probe's thermal transient, not the water.
  6. Immerse the probe fully, off all surfaces. Every sensing electrode submerged, and the probe not touching the sides or bottom of the container. Container walls distort the electrical field and produce false readings.
  7. Swirl gently while reading to displace the stagnant boundary layer at the electrode surface so the reading reflects bulk water composition.
  8. Record with full context — date, time, location, water temperature, EC in µS/cm, TDS in mg/L with the conversion factor used, and any observation about appearance or odor. A ppm value recorded without its factor is not reproducible.
  9. Rinse the probe with distilled water between samples to prevent carryover, particularly when moving from high-TDS to low-TDS samples.
  10. Apply the action threshold. NDSU's field trigger for laboratory submission is EC ≥ 6,000 µmhos/cm or TDS ≥ 4,500 ppm. Above that, field screening has done its job and speciated laboratory analysis is warranted.
Field errors that quietly corrupt data. Probe contacting the container wall or bottom. Testing in direct sun on a hot day without ATC. Reusing a container without rinsing between samples. Reading a NaCl standard before the salt has fully dissolved. Failing to recalibrate after a battery change. Sampling water with visible turbidity or foam — suspended solids and surfactants both perturb the reading, and turbid samples should be filtered before testing if suspended load is significant. None of these produce an obvious error; they all produce a plausible wrong number.

Interpreting the number

There is no universally good TDS value — only values appropriate to a use. The EPA's 500 mg/L figure is a secondary standard addressing taste, odor, and appearance, not a health-based enforceable limit, and it gets cited far outside the context it was written for.

TDS (mg/L)ClassificationApplication guidance
0–50Near-pure / ultrapureDistilled, deionized, RO permeate. Required for semiconductor, pharmaceutical, and reagent-grade applications. Aggressive toward metal at this purity — materials compatibility matters.
50–300ExcellentHigh-quality well water and RO product water. Suitable across food and beverage, aquaculture, hydroponics, and sensitive process applications.
300–600GoodSpans the EPA secondary standard of 500 mg/L. Acceptable for drinking water and most agricultural and industrial use.
600–900Acceptable — monitorTaste changes become noticeable in this band. Increase monitoring frequency; assess impact on sensitive processes and scale-forming equipment.
900–2,000MarginalAbove secondary drinking water guidance. Acceptable for many livestock classes. Boiler and cooling tower impact assessment warranted.
2,000–5,000High — restricted useNot suitable for drinking without treatment. Crops may show osmotic stress. Laboratory analysis recommended above 4,500 ppm per the NDSU threshold.
5,000–10,000Very high — treatment requiredUnsuitable for most livestock and most irrigation. Industrial use requires treatment. For scale: seawater is roughly 35,000 ppm.
Above 10,000BrineSpecialized industrial applications only. Substantial treatment required for any potable or agricultural use.
Compiled from EPA secondary drinking water standards, MSU Extension Water Quality Program, and NDSU Extension WQ1923. Application-specific standards override these general bands.

Two application-specific notes worth pulling out. In hydroponics, TDS isn't a contaminant measure at all — it's the nutrient dosing parameter, deliberately managed from around 400 ppm at seedling stage to 3,500+ ppm for heavy feeders. The 500 mg/L drinking water figure is irrelevant there, and the conversion-scale problem is acute because published crop targets rarely state which scale they assume. That's covered in detail in the hydroponics EC, TDS & pH guide. In RO systems, TDS is the performance metric: measure feed and permeate, and rejection percentage tells you membrane condition directly. Rising permeate TDS is the earliest available warning of membrane damage or O-ring bypass.

TDS is not hardness. Hardness is calcium and magnesium specifically. TDS is everything dissolved. Water can read 800 ppm TDS and be soft if the load is mostly sodium chloride, or read 350 ppm and be meaningfully hard if calcium and magnesium dominate. Neither converts to the other without ion-specific analysis — and a softener, which exchanges calcium and magnesium for sodium, removes hardness while leaving TDS essentially unchanged. If your softener didn't move the TDS reading, it's working correctly. See the hard water guide and the hardness calculator.

Livestock and agricultural thresholds

Agricultural operations are among the heaviest users of field TDS testing, and the thresholds are well established across the extension system. Elevated TDS reduces water intake, which reduces feed intake, which shows up as production loss before it shows up as an obvious health problem.

TDS (ppm)SuitabilitySpecies notes
Below 1,000Excellent for all classesNo concerns for any commonly tested livestock.
1,000–2,999Generally safeSatisfactory across species. May cause temporary mild diarrhea in unacclimated animals.
3,000–4,999Satisfactory for mostGenerally acceptable for cattle, horses, sheep, swine, poultry. Pregnant and lactating animals may be affected at the upper end — monitor intake.
5,000–6,999Unacceptable for poultry and swineReduced production in cattle and horses. Do not use for pigs, poultry, or young animals.
7,000–10,000Unacceptable for mostAvoid entirely for pregnant or lactating animals. Very limited use for dry beef cattle only.
Above 10,000Do not useSerious health risk for all livestock, including in drought emergency, without dilution or treatment.
Source: NDSU Extension WQ1923 and the broader US agricultural extension system. Species and acclimation status both modify these thresholds; consult a veterinarian for herd-specific decisions.

Sources of high TDS and treatment

TDS doesn't rise randomly. Each source has a characteristic signature, which is why the correct response to a high field reading is speciated laboratory analysis rather than immediate equipment purchase — the treatment depends on what is dissolved, not how much.

SourceContributesDiagnostic signature
Natural geologyCalcium, magnesium (limestone/dolomite); sulfate (gypsum); chloride (salt formations); iron and manganeseStable over time; matches regional aquifer chemistry. See the iron in well water guide for the iron case.
Agricultural runoff and irrigation returnNitrate, sulfate, sodium, chlorideElevated TDS with elevated nitrate. Seasonal correlation with application timing.
Road salt and deicersSodium chloride, calcium chloride, potassium acetateWinter/spring seasonal spike; sodium and chloride dominant; proximity to roads and lots.
Industrial effluentProcess-specific — metals, sulfates, dissolved organicsRequires speciated analysis. TDS indicates extent, never source.
Sewage or wastewater influenceNitrate, phosphate, dissolved organic matter, chlorideTDS with coliform and nutrient analysis for full characterization.
Distribution system corrosionIron, copper, lead; carbonates and bicarbonatesTDS rises between source and tap. Pair with a metals panel.
Boiler blowdown / cooling towerConcentrated feedwater minerals plus treatment chemistryTDS elevated versus upstream measurement; minerals match dosed chemistry.
TreatmentTDS reductionBest fitConsiderations
Reverse osmosis90–99%Drinking water, food and beverage, high-purity process water, aquaculture, hydroponicsMost versatile option. Produces a concentrate stream needing disposal. Requires pre-treatment for turbidity, iron, hardness, and chloramine — see the RO pre-treatment guide.
Distillation99%+Laboratory and pharmaceutical water; small-volume potableEnergy-intensive and slow. Impractical at volume. Output may need remineralization for palatability.
Electrodialysis (ED/EDR)50–90%Brackish desalination, larger-volume industrialWell suited to high-TDS feed. More complex than RO; uncommon at small commercial scale.
Ion exchangeVariable — ion-selectiveSoftening; deionization for ultrapure waterDoes not reduce total TDS in softening service — it substitutes sodium for calcium and magnesium. Full deionization is a different configuration.
NanofiltrationModerate; targets divalent ionsPartial softening; applications where full RO is excessivePasses monovalent ions. Lower energy than RO for equivalent flow.
BlendingProportional to ratioOperations with a second, lower-TDS supplyNo treatment capital cost. An EC meter is the essential instrument for verifying blend ratio continuously.
No treatment — monitor and document0%Sources below action thresholdsFrequently the correct answer. Regular measurement documents compliance and catches emerging trends early.

FAQ

What does a TDS meter actually measure?

Electrical conductivity, converted to a ppm estimate in firmware. The reference method for true TDS is gravimetric — evaporate a measured sample and weigh the residue, dried at 180 °C. Conductivity is a well-accepted proxy for that, not a direct measurement of it.

What is the EC to TDS conversion factor?

It varies with conductivity, from 0.50 at 100 µS/cm up to 0.82 at 10,000 µS/cm on the NDSU reference table. Meters that hard-code a single factor — commonly 0.50 for the NaCl scale or 0.70 for the 442 scale — are exact at one point and diverge from there, by as much as 39% at the extremes.

Why do two TDS meters give different readings on the same water?

Most often because they use different conversion scales, which alone accounts for up to a 40% spread between correctly functioning instruments. Other contributors are calibration state, temperature compensation differences, and cell constant mismatch with the sample range. Compare EC readings rather than ppm readings and most of the disagreement disappears.

What is a good TDS level in water?

Application-dependent. EPA's secondary standard is 500 mg/L for drinking water, an aesthetic rather than health-based limit, with under 300 mg/L generally preferred for taste. Livestock: under 3,000 ppm broadly safe. Hydroponics: 1,200–3,500 ppm by design. Ultrapure process water: under 10 ppm. A "good" number outside a stated application isn't meaningful.

Does a TDS meter measure water hardness?

No. Hardness is calcium and magnesium; TDS is all dissolved solids. High TDS with low hardness is common in sodium-dominated water. A properly working softener will barely change a TDS reading, because it exchanges hardness ions for sodium rather than removing dissolved mass.

Does "lab grade accuracy" mean anything on a TDS meter?

No. It's unregulated marketing language with no defining standard. The specifications that determine whether a reading is defensible are accuracy as a percentage of reading, cell constant, temperature compensation coefficient, resolution, and traceable calibration certification. If those aren't published, the adjective isn't a substitute.

How do I reduce TDS in water?

Dissolved solids can't be removed by conventional filtration, which only captures suspended particles. The options are reverse osmosis (most cost-effective in most commercial cases), distillation, electrodialysis, deionization, or blending with a lower-TDS supply. Which one is appropriate depends on what the dissolved load actually consists of, which is a laboratory question rather than a meter question.

When should I send a sample to a laboratory?

NDSU's field trigger is EC ≥ 6,000 µmhos/cm or TDS ≥ 4,500 ppm. Beyond that threshold, also send for compliance reporting, when establishing a baseline on a new source, when TDS has shifted more than about 20% from an established baseline, or whenever the number will be used by anyone outside your organization.

Related Guides and Reviews

Sources: Prasad, Laxmi, and Miranda Meehan. "Using Electrical Conductivity and Total Dissolved Solids Meters to Field Test Water Quality." NDSU Extension Publication WQ1923, North Dakota State University, revised January 2025. · Salverson, Robin. "Performing a Field Test for Livestock Water Quality." SDSU Extension, South Dakota State University, updated November 2023. · MSU Extension Water Quality Program / Montana Well Educated Program, "Alkalinity, pH, and Total Dissolved Solids," Montana State University. · US EPA National Secondary Drinking Water Regulations. · Standard Methods for the Examination of Water and Wastewater, Method 2540 C (Total Dissolved Solids Dried at 180 °C). Fixed-conversion-factor error figures in this guide were calculated from the NDSU WQ1923 reference table. This guide is informational and does not constitute engineering, regulatory, or veterinary advice.